Optical stimulation of the human eardrum for hearing devices – towards finite-element models for parameter optimization
Zusammenfassung:
Previous studies showed that irradiation of the eardrum with appropriately modulated pulsed laser light can lead to audible sound perception. Utilizing this phenomenon for hearing devices may provide benefits compared to traditional miniature loudspeakers, such as higher acoustic bandwidth without the need to seal the ear canal. Experiments with simple eardrum replica showed that the most efficient transduction mechanism is an optically-excited thermoelastic bending process, directly eliciting displacements of the eardrum or an attached passive material layer without primarily generating airborne sound. In this contribution, we show ongoing work towards establishing finite-element models of (artificial) human eardrums and middle ears, with the aim to systematically optimize transduction parameters. Initially, a simple model resembling the experimental conditions was established, and experimental results were reproduced with good accuracy. The results provide additional verification of the experimental outcomes, and yield detailed insights into the intricate relationships of the various physical mechanisms involved. First parameter studies were conducted using this model and revealed controllable factors influencing the transduction efficiency. Furthermore, optical stimulation was integrated into existing anatomical models of the middle ear, allowing to study the influence of more realistic geometries and boundary conditions.